Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
37 
3.8. Initial Calibration Procedure 
The next task is to calibrate the analyzer.   
To perform the following calibration you must have sources for zero air and span gas available for input 
into the sample port on the back of the analyzer.  See Section 3.1.3 for instructions for connecting these 
gas sources. 
While it is possible to perform this procedure with any range setting we recommend that you perform this 
initial checkout using the 500 ppm range.   
 
NOTE 
The following procedure assumes that the instrument does not have any of the available 
Zero/Span Valve Options installed. 
See Section 7.4 for instructions for calibrating instruments possessing Z/S valve options. 
 
1.  Set the Analog Output Range: 
 
 
 
SETUP X.X 
 
 CFG  DAS
  RNGE  
PASS  CLK  MORE  
EXIT
 
 SAMPLE          RANGE = 500.000 PPM        
CO2 
=X.XXX 
 < TST  TST >  CAL  
 
               SETUP 
 
 
 
SETUP X.X     
RANGE CONTROL MENU
 
 
 
 
 
MODE
 
 
 
 
 
 
SET  UNIT 
 
 
 
EXIT
 
Press this button to set 
the analyzer for  
 
 
 SNGL
 
 
 
DUAL 
 or AUTO ranges  
 
 
 
SETUP X.X 
RANGE: 500.000 CONC
 
 
 
 
 0          0         
 5
          
0          0         .
0
       ENTR   EXIT 
To change the value of the 
reporting range span, enter the 
number by pressing the key under 
each digit until the expected value 
appears. 
Press this button to select the 
concentration units of measure: 
 PPB, PPM, UGM, MGM
 
EXIT
 ignores the new setting and 
returns to the RANGE CONTROL 
MENU
. 
ENTR 
accepts the new setting and 
returns to the  
RANGE CONTROL MENU
 
 
 
SETUP X.X
RANGE: 
500.000 Conc
 
 
 
 0          0          0          5          0          .0     
 ENTR
   
EXIT