Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
41 
SAMPLE 
  GAS TO CAL:CO2
CO2
O2
  
ENTR EXIT
 SAMPLE       RANGE = 500.000 PPM      CO2 =XXX.X
 < TST  TST >      CAL                                        
SETUP
M-P CAL        A1:NXCNC1  =100PPM       NOX=X.XXX
 <TST TST> ZERO SPAN
CONC
EXIT
SAMPLE 
  GAS TO CAL:O2
NOX
O2
  ENTR
EXIT
M-P CAL          O2 SPAN CONC:20.8%
 0
2
0
.8
0
  ENTR EXIT
The O
2
 span concentration value automatically defaults to  
20.8 %
.  
If this is not the the concentration of the span gas being 
used, toggle these buttons to set the correct concentration 
of the O
2
 calibration gases.
EXIT
 ignores the new 
setting and returns to 
the previous display.
ENTR 
accepts the new 
setting and returns to 
the previous menu.
 
 
 
STEP 2 – ACTIVATE O
2
 SENSOR STABILITY FUNCTION 
To change the stability test function from NO
x
 concentration to the O
2
 sensor output, press: