Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
42 
SETUP X.X            
 STABIL_GAS:O2
  CO2   O2
 ENTR
EXIT
Press EXIT 
times to return 
to SAMPLE 
menu
SAMPLE       RANGE = 500.000 PPM       CO2 =XXX.X
 < TST  TST >      CAL                                       
 SETUP
SETUP X.X        
PRIMARY SETUP MENU
 CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X       
SECONDARY SETUP MENU
 COMM
VARS
DIAG ALRM
EXIT
  SETUP X.X         0) DAS_HOLD_OFF=15.0 Minutes
 <PREV
NEXT>
  JUMP
EDIT PRNT EXIT
 SETUP X.X         
2) STABIL_GAS=CO2
 <PREV  NEXT>  JUMP
EDIT
PRNT EXIT
SETUP X.X              
STABIL_GAS:CO2             
  CO2
O2
 ENTR EXIT
Continue pressing NEXT until ...
SETUP X.X      
ENTER PASSWORD:818
 
8
 
 
1
 
 
8
ENTR
EXIT
 
 
 
NOTE 
Use the same procedure to reset the STB test function to CO
2
 when the O
2
 
calibration procedure is complete.