Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
40 
3.8.1. Initial O
2
 Sensor Calibration Procedure 
If your instrument has an O
2
 sensor option installed that should be calibrated as well.   
3.8.1.1. O
2
 Calibration Setup 
The pneumatic connections for calibrating are as follows: 
C
a
li
bra
ted N
2
at
 HI
GH
 Span
 
Co
nc
ent
rat
ion
C
a
lib
ra
te
d
 O
2
at
 20.
8%
 Sp
an 
C
on
cen
tr
at
io
n
VE
NT
 
Figure 3-16: 
O2 Sensor Calibration Set Up 
O
2
 SENSOR ZERO GAS: Teledyne Instruments’ recommends using pure N
2
 when calibration the zero 
point of your O
2
 sensor option. 
O
2
 SENSOR SPAN GAS: Teledyne Instruments’ recommends using 21% O
2
 in N
2
 when calibration the 
span point of your O
2
 sensor option. 
 
3.8.1.2. O
2
 Calibration Method 
STEP 1 – SET O
2
 SPAN GAS CONCENTRATION : 
Set the expected O
2
 span gas concentration.   
This should be equal to the percent concentration of the O
2
 span gas of the selected reporting range 
(default factory setting = 20.8%; the approximate O
2
 content of ambient air).