Справочник Пользователя для Emerson Process Management Rosemount

Скачать
Страница из 4
Product Specification Sheet
00815-0100-4810, Rev AA
June 2005
Emerson Process Management
© 2005 Rosemount Inc. All rights reserved.
¢00815-0100-4810¤
Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc.
HART is a registered trademark of the HART Communications Foundation.
F
OUNDATION
 fieldbus is a registered trademark of the Fieldbus Foundation.
Hastelloy is a registered trademark of Haynes International, Inc.
Monel is a registered trademark of E.I. du Pont de Nemours & Co.
All other marks are the property of their respective owners.
Rosemount Inc.
8200 Market Boulevard
Chanhassen, MN 55317 USA
T (U.S.) 1-800-999-9307
T (International) (952) 906-8888
F (952) 949-7001
www.rosemount.com
Emerson Process 
Management GmbH & Co.
Argelsrieder Feld 3
82234 Wessling
Germany
T 49 (8153) 9390
F 49 (8153) 939172
Emerson Process Management Asia Pacific 
Private Limited
1 Pandan Crescent
Singapore 128461
T (65) 6777 8211
F (65) 6777 0947
Enquiries@AP.EmersonProcess.com
Mass Flow Flowmeters
For compressible fluid applications (i.e. steam and 
gases), multivariable differential pressure flowmeters 
are ideal. These flowmeters shall perform real-time 
calculations by compensating for changes in the 
COP flow coefficient (Cd) factor, gas expansion 
factor, velocity of approach factor, density or 
compressibility, viscosity, and Reynolds Number.
When a multivariable transmitter is integrated with 
the COP, the flowmeter shall perform within ±0.7% to 
±0.9% of mass flow rate in gas and steam. The 
vendor shall make independent testing 
documentation available demonstrating this 
performance.
Functional Specifications
The flowmeter shall provide a fully 
compensated mass flow output.
The flowmeter shall have the capability of 
either a Gage Pressure (GP) sensor or an 
Absolute Pressure (AP) sensor.
The pressure instrument shall be two-wire, 
multivariable (differential pressure, pressure, 
and temperature).
Software Specifications
Pressure Instrument outputs shall be a 4-20 
mA analog signal, user-selectable to represent 
mass flow, differential pressure, static 
pressure, or temperature, with a 
superimposed digital signal, using HART
®
 
protocol.
The pressure instrument shall perform 
continuous diagnostics, capable of self-test 
functions and be able to provide specific 
diagnostic information.
Basic configuration capabilities of the pressure 
instrument shall allow the user the ability to 
input and store information including the 
range, engineering units, damping, drain / vent 
valves, flange, and O-ring materials, date, 
message, descriptor, tag number, and serial 
number.
Mass flow configuration shall allow the user to 
input and store information including the fluid 
name, fluid density, fluid viscosity and the 
COP primary element calibration factor.
To calculate mass flow, the transmitter shall utilize 
the full DP mass flow equation:
Where, 
Q
m
 = Mass Flow
N = Units Conversion Factor
C
d
 = Discharge Coefficient
Y
1
 = Gas Expansion Factor
E = Velocity of Approach Factor
d
2
 = Bore of the Differential Producer
ρ=Density
DP = Differential Pressure
Q
m
NC
d
EY
1
d
2
DP ρ
( )
{
}
1
2
-
=