Справочник Пользователя для Emerson 5081-P

Скачать
Страница из 156
86
MODEL 5081-P pH/ORP
SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
8.6 DISPLAy UNITS
8.6.1 Purpose
This section describes how to do the following:
1. switch the process display units between pH and ORP (millivolts),
2. select °C or °F for the temperature display,
3. select percent of full scale or milliamps for the output display,
4. program a security code.
8.6.2 Definitions
1. dISPLAY  uNITS.  Select  pH  if  the  transmitter  is  being  used  to  measure  pH.  Select
ORP  if  the  transmitter  is  being  used  to  measure  ORP.  ORP  is  oxidation-reduction
potential. ORP has units of millivolts and is usually measured with an inert metal elec-
trode, such as a platinum electrode. The units selected are shown in the main display
next to the measured value.
2. OuTPuT CuRRENT dISPLAY (5081-P-HT only). The transmitter generates a 4 to 20
mA  output  signal  directly  proportional  to  the  pH  of  the  sample.  The  output  signal
appears on the same line with the temperature. The output signal can be displayed
as current (in mA) or as percent of full scale.
3. SECuRITY  COdE. The  security  code  unlocks  the  transmitter  and  allows  complete
access to all menus. The transmitter is shipped with the security code disabled.
8.6.3 Procedure
1. Press  PROG  on the infrared remote controller (IRC).
2. Press  NEXT  until the dISPLAy sub-menu appears. Press  ENTER .
3. The screen displays the tyPE prompt. Press é or ê to toggle between pH and OrP.
Press  ENTER  to save.
4. The  screen  displays  the tEMP prompt.  Press é or  ê to toggle between and F.
Press  ENTER  to save.
5. The  screen  displays  the  OUtPUt  prompt.  Press é or ê to toggle between and
CUR.  Press  ENTER  to save.
6. The screen displays the COdE prompt. use the editing keys to enter a security code
between 001 and 999. Entering 000 disables the security feature. Press  ENTER  to
save. The security code does not become effective until about two minutes after the
last keystroke.
7. Press  EXIT  to return to the process display.
PROGRAM
OUtPUt
EXIT
ENTER
CUR
PROGRAM
CODE
EXIT
ENTER
0
00
PROGRAM
tYPE
EXIT
ENTER
PH
PROGRAM
tEMP
EXIT
ENTER
C
PROGRAM
dISPLAY
EXIT
NEXT       ENTER