MODEL 5081-P pH/ORP
SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
3. STABILITY CRITERIA. For the Model 5081-P pH/ORP transmitter to accept calibra-
tion data, the pH must remain within a specified range for a specified period of time. The
default values are 0.02 pH units and 10 seconds. In other words, at the default setting,
calibration data will be accepted as soon as the pH reading is constant to within 0.02 units
for 10 seconds. The minimum range is 0.01, and the maximum time is 30 seconds.
8.7.3 Procedure
1. Press  PROG  on the infrared remote controller (IRC).
2. Press  NEXT  until the bUFFEr sub-menu appears. Press  ENTER .
3. The screen displays the bAUTO prompt. Press é or ê to activate (On) or deacti-vate
(OFF) auto calibration. Press  ENTER  to save.
4. The screen displays the bUFFEr prompt. Press é or ê to select the desired buffer or
buffers. The buffer values available under each designation are given in Table 8-4.
Press  ENTER  to save.
TABLE 8-4. Standard and Technical Buffers Recognized by the 5081pH Transmitter
Std
NIST, dIN 19266, JIS 8802, and BSM standard buffers
ErC
Merck Buffers (technical buffers)
InG
Ingold Buffers (technical buffers)
din
dIN 19267 (technical buffers)
FSH
Fisher Buffers (technical buffers)
5. The screen changes to display the tIME prompt. use the editing keys to change the
flashing number to the time in seconds the reading must remain stable for calibration
data to be accepted. The maximum is 30 seconds. Press  ENTER  to save.
6. The screen changes to display the pH prompt. use the editing keys to change the
flashing display to the pH range the reading must remain in for calibration data to be
accepted. The minimum range is 0.01. Press  ENTER  to save.
7. Press  EXIT  twice to return to the process display.
Merck 
Ingold  
DIN 19267  
Fisher
pH
temp (°C)
pH
temp (°C)
pH
temp (°C)
pH
temp (°C)
1.09
0 - 90
1.0
0 - 60
2.00
0 - 95
2.00
0 - 95
2.0
0 - 60
3.06
0 - 90
3.0
0 - 60
4.01
0 - 95
4.0
0 - 60
4.65
0 - 90
5.0
0 - 60
6.79
0 - 90
6.0
0 - 60
7.00
0 - 95
7.00
0 - 95
7.0
0 - 60
8.0
0 - 60
9.00
0 - 95
9.21
0 - 95
9.23
0 - 90
8.99
0 - 60
10.0
0 - 60
11.0
0 - 60
12.00
0 - 95
12.75
0 - 90
TABLE 8-3. pH values of commercial (technical) buffers and the temperature range over which pH values are defined
PROGRAM
bAUtO
EXIT
ENTER
On
PROGRAM
bUFFEr
EXIT
ENTER
Std
PROGRAM
bUFFEr
EXIT
NEXT       ENTER
PROGRAM
tIME
EXIT
ENTER
1
0
PROGRAM
pH
EXIT
ENTER
0
0
.02