SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
8.7 BUFFER CALIBRATION PARAMETERS
8.7.1 Purpose
This section describes how to do the following:
1. activate or deactivate auto calibration,
2. identify which buffers will be used during auto calibration,
3. set the stability criteria for auto calibration.
8.7.2 Definitions
1. AuTO CALIBRATION. In auto calibration, screen prompts direct the user through a two point buffer calibration. The
transmitter recognizes the buffers and uses temperature-corrected values in the calibration. The transmitter does not
accept data until programmed stability limits have been met. If auto calibration is deactivated, the user must perform
a manual calibration. In manual calibration, the user judges when readings are stable and manually enters buffer val-
ues. The use of auto calibration is strongly recommended.
2. BuFFERS.  Buffers  are  aqueous  solutions  to  which  exactly  known  pH  values  have  been  assigned. Assigning  a  pH
value to a buffer involves certain fundamental assumptions. Slightly different assumptions lead to slightly different pH
scales. Over the years, various national standards organizations have developed different scales. The Model 5081-P
pH/ORP  transmitter  recognizes  the  common  standard  scales  as  well  as  common  commercial  buffers.  Commercial
buffers, which are sometimes called technical buffers, are traceable to standard buffers, but the accuracy of commer-
cial buffers is generally less than standard buffers. Tables 8-2 and 8-3 list the buffers the Model 5081-P pH/ORP trans-
mitter recognizes and the temperature range over which the buffer pH is defined.
NOTE:  pH 7 buffer is not a standard buffer. Because it is a popular commercial buffer in the united States, it is included
with the standard buffers. The pH of the buffer is defined between 0 and 95°C.
NIST  
DIN 19266  
JIS 8802  
BSI
pH
temp (°C)
pH
temp (°C)
pH
temp (°C)
pH
temp (°C)
1.68
5 - 95
1.68
5 - 95
1.68
0 - 95
1.68
0 - 60
3.56
25 - 95
3.56
25 - 60
3.78
0 - 95
4.01
0 -95
4.01
0 - 95
4.01
0 - 95
4.01
0 - 60
6.86
0 -95
6.86
0 - 95
6.86
0 - 95
6.86
0 - 60
7.00
see note
7.41
0 - 50
9.18
0 - 95
9.18
0 - 95
9.18
0 - 95
9.18
0 - 60
10.01
0 - 50
10.01
0 - 50
10.01
0 - 50
12.45
0 - 60
12.45
0 - 60
TABLE 8-2. pH values of standard buffer solutions and the temperature range over which pH values are defined