3M™ Rapid Detection Reader 
Troubleshooting 
39 
Power On Self Test Message / IQC Messages 
Message 
Cause 
Action 
No Available Ports. IQC Failed 
To Start.
 
Ports are disabled. 
All ports are disabled.  Switch 
the CM off and switch back on.  
If problem persists contact 
Technical Support to report the 
problem. 
Test Run Messages 
To display the error message, before removing the cartridge, touch the red port icon for 
detailed information about the failure.  Once the cartridge is removed, this information is sent to 
the appropriate Log files.  Refer to Viewing Saved Logs for additional information. 
Test Run Messages
 – indicated by red port icon 
 
 
Port Display 
Error Message 
Cause (s) 
Action 
CTG 
Cartridge Error 
Possible defect in Test Cartridge. 
HRB 
High Background 
Possible interfering substances 
in the sample causing high 
running background. 
LOW 
Low Signal 
Insufficient sample transferred to 
the Test Cartridge. 
Insufficient mixing of the sample 
with the supplied assay tip. 
Interfering substances in the 
sample. 
Fail 
Sample Error 1 
Sample flow occurred before the 
Test Cartridge was inserted into 
the port.  
Delay of more than 30 seconds 
between sample addition and 
Test Cartridge insertion. 
Insertion of a used Test 
Cartridge. 
The Test Cartridge is 
ejected. 
Remove and dispose of 
appropriately. 
Repeat the assay with a 
fresh set of components. 
If problem persists 
contact Technical 
Support to report the 
problem.