A blue stripe on the label identifies the SCC-TC0X as an analog input 
module. The label also displays the icon shown in Figure 1.
Figure 1.  SCC-TC0X Icon
You can plug the SCC-TC0X into any analog input socket on the SC-2345. 
The socket you select determines which E Series DAQ device channels 
receive the SCC-TC0X signals.
For single-stage input conditioning, plug the SCC-TC0X into any socket J 
(X +1), where X is 0 to 7, and connect the input signals to the module as 
described in the Connecting the Input Signals section.
If you use the SCC-TC0X in a dual-stage configuration, the SCC-TC0X 
must be the first-stage module. Plug it into any socket J (X+9) and plug the 
second-stage SCC into socket J (X+1), where X is 0 to 7. Connect the input 
signals to SCC-TC0X as described in the Connecting the Input Signals 
section. The SC-2345 connects the output signals of the first-stage SCC to 
the inputs of the second-stage SCC. An example of dual-stage conditioning 
in an SCC-TC0X followed by an SCC-LP01 lowpass filter module.
Sockets J9 to J16 also are available for digital input/output (DIO) 
conditioning or control. Refer to the SC-2345 User Manual for more 
information on configuring, connecting and installing SCC modules
Note
SCC-TC0X dual-stage configuration is only available in NI-DAQ 7.1 or later.
Connecting the Input Signals
The SCC-TC01 has a two-prong uncompensated thermocouple 
miniconnector that accepts any miniature or subminiature two-prong 
male thermocouple connector. The SCC-TC02 has a three-position 
screw-terminal connector that accepts 28 to 16 AWG thermocouple wires. 
The screw-terminal connector provides a ground connection for shielded 
thermocouples. Otherwise, the two modules function identically.
To use an SCC-TC01, plug the male thermocouple connector into the 
module.
TC