SCC-TC Series Thermocouple Input Modules User Guide
The SCC-TC02 has a fixed screw-terminal receptacle and a removable 
screw-terminal block, as shown in Figure 2.
Figure 2.  SCC-TC02 Two-Part Screw-Terminal System
After you install the SCC-TC02, attach the input signals to the screw 
terminals of the module.
1.
Remove power from the signal lines.
2.
Strip 7 mm (0.28 in.) of insulation from the ends of the signal wires.
3.
Insert the wires into the screw terminals.
4.
Tighten the screws to 0.5 to 0.6 N
⋅ m (4.4 to 5.3 lb −  in.) of torque.
The SCC-TC02 accepts up to three signals: TC
+, TC–, and GND. TC+ is 
the positive thermocouple lead and TC– is the negative thermocouple lead.
1
 
The GND terminal connects to AIGND on the E Series DAQ device.
The SCC-TC0X has a 10 M
Ω bias resistor connected from the negative 
thermocouple input to ground. This resistor allows the thermocouple to be 
ground-referenced or floating without requiring external bias resistors 
connected to ground.
1
SCC Screw-Terminal Receptacle
2
Removable Screw-Terminal Block
1   
For ANSI color-coded J-type thermocouples, the red wire is negative and the white wire is positive. Refer to the thermocouple 
data sheet if possible. You can find information about other color-coding schemes in the NI KnowledgeBase at 
ni.com/
support
3