BAROSTIM NEO LEGACY REFERENCE GUIDE
4-3
Carotid Sinus Lead
Warnings
• Patients who manipulate the Carotid Sinus Lead through the skin may damage or disconnect the lead
from the IPG and/or possibly cause damage to the carotid sinus.
• Lead malfunction could cause painful stimulation and/or stimulation of adjacent tissue.
Precautions
• Electrocautery at a low but effective power can be used to minimize the potential of damaging the leads
during dissection. Electrocautery at high power settings may damage the Carotid Sinus Lead.
• Scalpels may damage the Carotid Sinus Lead. Avoid scalpel blade contact with the lead when using
scalpels.
• Exercise extreme caution in utilizing line-powered equipment in conjunction with the Carotid Sinus Lead
because leakage current could injure the patient.
CVRx Programmer System
Warning
• Do not locate any programmer system components inside the sterile operating field.
Precautions
• The components of the Programmer System should not be sterilized.
• The following are requirements to comply with IEC 60601-1 and IEC 60601-1-1:
• The following are requirements to comply with IEC 60601-1 and IEC 60601-1-1:
• The computer and power supply should be located outside the patient environment when the
computer is operated on mains power.
• The system should not be connected to other non-isolated monitoring equipment or
communication networks.
• The operator should not touch the computer and the patient simultaneously when the
computer is operated on mains power.
• The USB cable should be fully inserted into the Programmer Interface USB receptacle to
avoid patient contact with the metal part of the USB connector.
Note: The patient environment is defined as the area within 1.5m (approximately 5ft) of the
patient.
patient.
• Plug the Programmer System directly into an outlet or operate using the laptop battery power. Do not
plug the programmer system into a power strip.
• Do not immerse product in water or a safety hazard could arise during use. For cleaning instructions,
refer to Section 7.
• Keep the Programmer System in a controlled location to prevent loss or theft. Intentional misuse of the
Programmer System could result in an IPG being programmed to settings that are not as prescribed.