How Supplied
The implantable components of the Neo Legacy system and implant accessories have been sterilized using
ethylene oxide gas. CVRx has supplied these components in a sterile package for direct introduction into the
operating field.
ethylene oxide gas. CVRx has supplied these components in a sterile package for direct introduction into the
operating field.
Implantable Pulse Generator
Supplied as one sterile IPG Model 2100 with port plug and torque wrench. When the device is
shipped, the therapy schedule is OFF.
shipped, the therapy schedule is OFF.
Programmer System
Programmer Interface Model 9010 with USB cable.
The Programmer Software is to be installed on a computer equipped with a USB interface.
Inspection Prior to Use
Implantable Pulse Generator
Carefully inspect the IPG sterile package before opening.
The implantable components are supplied STERILE and for SINGLE USE
Do not use on or after the “Use By” date. Return the unopened package to CVRx.
. Do not use if the
package is opened or damaged. Return the package and/or contents to CVRx. Reuse of this product
may result in malfunction or adverse events such as infection or death.
may result in malfunction or adverse events such as infection or death.
Before opening the IPG package, establish a communication session with the IPG. If the reported
battery voltage is less than 2.85V, return the package, unopened, to CVRx.
battery voltage is less than 2.85V, return the package, unopened, to CVRx.
Materials Recommended for Implantation and/or Explantation
• A table or stand outside the sterile operating field to hold the Programmer System.
• Optional Accessories
• Optional Accessories
- CVRx Accessory Kit, Model 5500
Cleaning the Programmer System
If the Programmer System requires cleaning, clean the system components with a soft cloth dampened with
water. Do not allow pooling or ingress of liquid into the Programmer Computer or the Programmer Interface
enclosure.