BAROSTIM NEO LEGACY REFERENCE GUIDE 
4-3
 
Carotid Sinus Lead 
Warnings 
•  Patients who manipulate the Carotid Sinus Lead through the skin may damage or disconnect the lead 
from the IPG and/or possibly cause damage to the carotid sinus. 
•  Lead malfunction could cause painful stimulation and/or stimulation of adjacent tissue. 
Precautions 
•  Electrocautery at a low but effective power can be used to minimize the potential of damaging the leads 
during dissection.  Electrocautery at high power settings may damage the Carotid Sinus Lead. 
•  Scalpels may damage the Carotid Sinus Lead.  Avoid scalpel blade contact with the lead when using 
scalpels. 
•  Exercise extreme caution in utilizing line-powered equipment in conjunction with the Carotid Sinus Lead 
because leakage current could injure the patient. 
CVRx Programmer System 
Warning 
•  Do not locate any programmer system components inside the sterile operating field. 
Precautions 
•  The components of the Programmer System should not be sterilized. 
•  The following are requirements to comply with IEC 60601-1 and IEC 60601-1-1: 
•  The computer and power supply should be located outside the patient environment when the 
computer is operated on mains power. 
•  The system should not be connected to other non-isolated monitoring equipment or 
communication networks.   
•  The operator should not touch the computer and the patient simultaneously when  the 
computer is operated on mains power.  
•  The USB cable should be fully inserted into the Programmer Interface USB receptacle to 
avoid patient contact with the metal part of the USB connector.   
Note: The patient environment is defined as the area within 1.5m (approximately 5ft) of the 
patient.    
•  Plug the Programmer System directly into an outlet or operate using the laptop battery power. Do not 
plug the programmer system into a power strip. 
•  Do not immerse product in water or a safety hazard could arise during use.   For cleaning instructions, 
refer to Section 7. 
•  Keep the Programmer System in a controlled location to prevent loss or theft.  Intentional misuse of the 
Programmer System could result in an IPG being programmed to settings that are not as prescribed.