How Supplied 
The implantable components of the Neo Legacy  system  and implant accessories have been sterilized using 
ethylene oxide gas. CVRx  has supplied these components in a sterile package for direct introduction into the 
operating field. 
Implantable Pulse Generator 
Supplied as one  sterile IPG  Model 2100  with port plug and torque wrench.  When the device is 
shipped, the therapy schedule is OFF
Programmer System 
Programmer Interface Model 9010 with USB cable. 
The Programmer Software is to be installed on a computer equipped with a USB interface. 
Inspection Prior to Use 
Implantable Pulse Generator 
Carefully inspect the IPG sterile package before opening. 
The implantable components are supplied STERILE  and for SINGLE USE
Do not use on or after the “Use By” date. Return the unopened package to CVRx. 
.  Do not use if the 
package is opened or damaged. Return the package and/or contents to CVRx.  Reuse of this product 
may result in malfunction or adverse events such as infection or death. 
Before opening the IPG package,  establish a communication session with the IPG. If the reported 
battery voltage is less than 2.85V, return the package, unopened, to CVRx. 
Materials Recommended for Implantation and/or Explantation 
•  A table or stand outside the sterile operating field to hold the Programmer System. 
•  Optional Accessories 
-  CVRx Accessory Kit, Model 5500 
Cleaning the Programmer System 
If the Programmer System requires cleaning,  clean the system components with a soft cloth dampened with 
water. Do not allow pooling or ingress of liquid into the Programmer Computer or the Programmer Interface 
enclosure.