Temperature 
Between  -20°C  à  35°C  (For  6months 
maximum) 
45°C (For 1 month maximum) 
Shock resistance 
4600 g for a time of 0,5 ms 
Magnetic field survivability 
10 000 Gauss 
Maximum  rate  of  magnetic  field  tolerated  before  miss-
calibration 
 
20 Gauss 
 
MPOD  transmits  data  to  the  CTRLER via  radio  signal. The transmitted signal  is in 
the Bluetooth band. MPOD is tested in CEM. However, the presence of other radio 
transmitting  devices  can  lead  to  interferences  with  MPOD™.  Do  not  to  use  the 
MPOD  when  other  radio  operating  devices  are  running  and  are  priority  for 
patients. 
 
 
 
The  device  is  mobile  and  portable.  Its  transporting  conditions  must  respect  the 
intended  protection  means.  Internal  transportation  does  not  require  particular 
precautions 
IX.
 
CALIBRATION 
When to calibrate? 
Your MPOD has been calibrated during the manufacturing process. Calibration consists in adjusting 
certain  parameters  for  offset  and  sensitivity  and  to  obtain  standardized  measurements.  However, 
over  a  long  period  of  time,  or  after  a  magnetic  shock,  the  sensors  are  likely  to  drift.  It  is  then 
necessary to calibrate. 
In static mode,  irregularities on accelerometer measures can be  due to a very  fast movement.  The 
software  will  probably  request  re-recording  with  a  slower  gesture.  If  magnetometers  are 
responsible  for  the  anomaly,  it’s  more  likely  because  there  are  magnetic  disturbances  in  the 
environment. In all other cases, it is necessary to calibrate. 
It is recommended to calibrate MPOD manually before any procedure. Calibration lasts 30 seconds 
and  it  is  well  guided  by  the  provided  MotionDevTool  v2.0.4  Software.  It  mainly  performs 
measurements, and has menus to calibrate the MPOD according to the procedure described below. 
How to calibrate? 
Here’s the best way to calibrate the MPOD: 
 
Be  sure  that  the  zone  where  the  calibration  is  carried-out  is  free  from  any  magnetic 
disturbances (refer to section VII). 
 
Take the sensor only in hand. We  recommend putting  your elbows on  a wooden  table  and 
maintaining the MPOD with your fingers 
 
It is advised to be oriented facing north.  
 
Rotate the MPOD with your fingers according to the 3 rotation axes (figure below):