Справочник Пользователя для Mindray DS USA Inc. WT0707-00000
Panorama™ Operating Instructions
0070-10-0690-01
5 - 25
Patient Setup Functions
Transfer Tab
2. Ensure that the device name for the Source Panorama is properly entered into the
Equipment List at the Destination Panorama, as described in the ‘‘Equipment Setup Tab’’
on page 9-33.
on page 9-33.
3. Ensure that the device name for the Destination Panorama is properly entered into the
Equipment List at the Source Panorama, as described in the ‘‘Equipment Setup Tab’’ on
page 9-33.
page 9-33.
4. Select the patient tile to be transferred. Choose a patient tile by selecting its digital data
or waveform area.
5. Select the Central To Central Without Device button. The Select Destination
FIGURE 5-15 Select Destination Panorama List Box
6. Select the Destination Panorama device from the Select Destination Panorama list
box.
• Select the Done button. The message Are you sure you want to transfer the
selected patient? is displayed with the following two choices:
• Select the Cancel button to terminate the transfer and return to the Transfer tab.
If the transfer process is initiated, it will continue as described in following section,
“Destination Panorama”. As the transfer process continues, the procedure will refer to
messages and a progress bar that will display at the Source Panorama.
“Destination Panorama”. As the transfer process continues, the procedure will refer to
messages and a progress bar that will display at the Source Panorama.
5.2.3.2
Destination Panorama
When a Central To Central Transfer Without Device has been requested from
another Panorama Central Station, the message Transfer In Request from [the Source
Panorama name]. Accept/Reject in Transfer Dialog. is displayed in the system
status line of the Destination Panorama.
another Panorama Central Station, the message Transfer In Request from [the Source
Panorama name]. Accept/Reject in Transfer Dialog. is displayed in the system
status line of the Destination Panorama.
1. At the Destination Panorama, select a tile that has a monitoring device attached.
2. From the menu bar, select the Patient Setup menu button. The tabs associated with
Patient Setup are displayed.
• Select the Yes button to initiate the first part of the transfer process and
return to the Transfer tab. In the Transfer tab, the message Central
Transfer in progress is displayed next to the Central To Central
Without Device button, which has been disabled.
Transfer in progress is displayed next to the Central To Central
Without Device button, which has been disabled.
• Select the No button to terminate the transfer and return to the Transfer
tab.