11-10
VC150 Vital Signs Monitor
KO00065K
Welch Allyn temperature: Welch Allyn temperature calibration and self-checks
6. Separate the buttocks with one gloved hand and gently insert the probe tip 
according to hospital protocol – but no further than 1.5 cm (0.6 inches) for 
adults (less for infants and children).
WARNINGS
If the tip is inserted too far, patient tissue damage may occur and the 
probe tip may not have good contact with tissue. Use of a lubricant on 
the probe cover is optional.
Before performing rectal temperature measurement on neonates and 
children, check hospital policy whether it is allowed and with what 
conditions.
7. Always hold the probe in place, maintaining tissue contact until temperature 
is complete. Do not allow the patient to reposition the probe.
8. When the predictive measurement is complete, the result is captured to 
patient data automatically. If you question a reading, select the snail icon to 
switch to monitor mode.
9. When the monitor mode temperature measurement has stabilized, select 
Snapshot to create a snapshot and remove the probe from the patient.
10. Eject the disposable probe cover by firmly pressing the ejection button on 
the top of the probe. Then remove gloves.
CAUTION
To prevent cross-contamination, properly dispose of the probe cover 
and gloves when the measurement is complete.
11. Place the probe in the probe well. Once you place the probe in the probe 
well, the temperature values will be cleared in 1 to 30 minutes, depending 
on the measurement expiration time setting.
Welch Allyn temperature calibration and self-checks
When the monitor is powered on, the monitor automatically calibrates the 
temperature circuit to account for ambient room temperature.
NOTE
If large changes occur in the ambient temperature, the temperature system 
displays a technical error message. Turn the unit power off and on again. 
Note that the patient will be discharged and non-default settings cleared 
when the monitor is turned off. If the unit is not turned off after this 
temperature error message, the temperature measurement may not be 
accurate.