11-9
Welch Allyn temperature: Welch Allyn temperature measurement
13. Eject the disposable probe cover by firmly pressing the ejection button on
the top of the probe.
CAUTION
To prevent cross-contamination, properly dispose of the probe cover
when the measurement is complete.
when the measurement is complete.
14. Place the probe in the probe well. Once you place the probe in the probe
well, the temperature values will be cleared in 1 to 30 minutes, depending
on the measurement expiration time setting.
on the measurement expiration time setting.
Rectal temperature measurement
Use gloves or other measures to reduce the risk of cross-contamination as
appropriate for good clinical practice and/or your institution's infection control
policies.
appropriate for good clinical practice and/or your institution's infection control
policies.
1. Check that a red probe well and a probe with red ejection button are
connected to the device. If not, obtain a red probe well and a probe with the
red ejection button and connect these to the device as instructed in
red ejection button and connect these to the device as instructed in
on page 3-4. If these are already connected,
ensure that the probe is in the probe well.
2. Always place a new transparent protective temperature probe cover on the
probe before every rectal use:
•
Remove the probe from the probe well.
•
Insert the probe shaft into a probe cover and firmly press down until
retaining rings of the probe cover seat securely over the retaining ring
barb of the probe shaft.
retaining rings of the probe cover seat securely over the retaining ring
barb of the probe shaft.
3. Verify that the probe cover fits snugly.
NOTES
Failure to firmly install the probe cover may result in the probe cover
becoming loose or disengaged during use. Unintended probe cover
ejection can lead to patient injury.
becoming loose or disengaged during use. Unintended probe cover
ejection can lead to patient injury.
Do not insert the probe cover into the probe well.
Do not allow the probe tip to come into contact with the patient until the
probe is placed in the measurement site. Before this, any contact
between the probe tip and the tissue or other material may cause
inaccurate readings.
probe is placed in the measurement site. Before this, any contact
between the probe tip and the tissue or other material may cause
inaccurate readings.
CAUTIONS
Injury may occur as a result of patient movement during the procedure.
Do not insert the probe cover into the probe well.
4. Select the temperature parameter to activate a shortcut to the Monitor
Setup > Temperature screen.
5. Select Measurement Site. Check that the temperature area displays Rectal.
If this is not displayed, reselect the measurement site.