11-8
VC150 Vital Signs Monitor
KO00065K
Welch Allyn temperature: Welch Allyn temperature measurement
4. Verify that the probe cover fits snugly.
NOTES
Failure to firmly install the probe cover may result in the probe cover 
becoming loose or disengaged during use. Unintended probe cover 
ejection can lead to patient injury.
Do not allow the probe tip to come into contact with the patient until the 
probe is placed in the measurement site. Before this, any contact 
between the probe tip and the tissue or other material may cause 
inaccurate readings.
CAUTIONS
Injury may occur as a result of patient movement during the procedure.
Do not insert the probe cover into the probe well.
5. Select the temperature parameter to activate a shortcut to the Monitor 
Setup > Temperature screen.
6. Select the adult or pediatric axillary icon on the Measurement Site screen.
7. Avoid folds in the axilla and place the probe tip as high as possible in the 
mid-axilla. Left image below is for adults, right image for infants and 
children.
 
8. Insert the probe in the patient's axilla, making 
sure the tip of the probe is in contact with the 
skin and positioned as close as possible to the 
axillary artery.
9. Move the patient’s arm close to the body to seal 
off the probe in the armpit. Keep the arm from 
moving during the measurement cycle.
10. Always hold the probe in place, maintaining tissue contact until temperature 
is complete. Do not allow the patient to reposition the probe.
11. When the predictive measurement is complete, the result is captured to 
patient data automatically. If you question a reading, select the snail icon to 
switch to monitor mode.
12. When the monitor mode temperature measurement has stabilized, select 
Snapshot to create a snapshot and remove the probe from the patient.
Adult
Pediatric
axillary
axillary
Oral