Welch Allyn temperature: Welch Allyn temperature measurement
4. Verify that the probe cover fits snugly.
NOTES
Failure to firmly install the probe cover may result in the probe cover
becoming loose or disengaged during use. Unintended probe cover
ejection can lead to patient injury.
becoming loose or disengaged during use. Unintended probe cover
ejection can lead to patient injury.
Do not allow the probe tip to come into contact with the patient until the
probe is placed in the measurement site. Before this, any contact
between the probe tip and the tissue or other material may cause
inaccurate readings.
probe is placed in the measurement site. Before this, any contact
between the probe tip and the tissue or other material may cause
inaccurate readings.
CAUTIONS
Injury may occur as a result of patient movement during the procedure.
Do not insert the probe cover into the probe well.
5. Select the temperature parameter to activate a shortcut to the Monitor
Setup > Temperature screen.
6. Select the adult or pediatric axillary icon on the Measurement Site screen.
7. Avoid folds in the axilla and place the probe tip as high as possible in the
mid-axilla. Left image below is for adults, right image for infants and
children.
children.
8. Insert the probe in the patient's axilla, making
sure the tip of the probe is in contact with the
skin and positioned as close as possible to the
axillary artery.
skin and positioned as close as possible to the
axillary artery.
9. Move the patient’s arm close to the body to seal
off the probe in the armpit. Keep the arm from
moving during the measurement cycle.
moving during the measurement cycle.
10. Always hold the probe in place, maintaining tissue contact until temperature
is complete. Do not allow the patient to reposition the probe.
11. When the predictive measurement is complete, the result is captured to
patient data automatically. If you question a reading, select the snail icon to
switch to monitor mode.
switch to monitor mode.
12. When the monitor mode temperature measurement has stabilized, select
Snapshot to create a snapshot and remove the probe from the patient.
Adult
Pediatric
axillary
axillary
Oral