Справочник Пользователя для Innokas Yhtyma Oy VC150
KO00065K
VC150 Vital Signs Monitor
D-11
Supplemental analysis of clinical accuracy test data: Masimo sensor accuracy
M-LNCS/LNCS/LNOP - Inf/Neo/NeoPt
M-LNCS/LNCS/LNOP - Inf/Neo/NeoPt, measured values
Range
Arms
90-100%
1.85%
80-90%
1.44%
70-80%
0.89%
Overall Claimed Accuracy Value
Range
Arms
Inf
Neo*
Neo Pt*
70-100%
± 2%
± 2% Adult
± 3% Neonatal
± 3%
*The saturation accuracy of the Neonate and Preterm sensors was validated
on adult volunteers and 1% was added to account for the properties of fetal
hemoglobin.
on adult volunteers and 1% was added to account for the properties of fetal
hemoglobin.
SaO
2
Err
or (SpO
2
- S
aO
2
)