Performance of Measurements
6.3 Trouble-Shooting
The UNITEST Echometer 3000 enables to locate
faults in cables or conductors.
The UNITEST Echometer 3000 enables to locate
faults in cables or conductors.
The user must check that the circuits are not
live !
live !
" Switch on the UNITEST Echometer 3000 press-
ing the
key.
" Change to the reference range pressing the
key, e.g. Reference level I. REF I is in-
dicated on the display.
" Connect the red test lead to the red socket and
the black test lead to the black socket of the m
range.
range.
" Connect the faulty leads of the conductor to be
measured using the alligator clamps.
" Set the measurement unit (m/ft) pressing the
key. ''m'' or ''ft'' is displayed.
" Set the measurement range selection dial in ac-
cordance with the conductor types in Tables I or
II. The
II. The
key is used to change to the
measurement ranges of Table II. ''TAB II'' is in-
dicated on the display.
dicated on the display.
If the message ''- - -'' is displayed, the UNITEST
Echometer is ready to perform measurements.
Echometer is ready to perform measurements.
" Start the measurement pressing the
key.
A single signal sound is audible and the mea-
surement value of the conductor length in the
pre-selected unit (m/ft) up to the error location
is displayed. If the message "COMP is dis-
played, the leads have a short-circuit.
surement value of the conductor length in the
pre-selected unit (m/ft) up to the error location
is displayed. If the message "COMP is dis-
played, the leads have a short-circuit.
Short-circuited cable ends cause an increasing
measurement error of the UNITEST Echometer
3000.
measurement error of the UNITEST Echometer
3000.
6.4 Performance of Resistance Measurements
The UNITEST Echometer 3000 can easily and pre-
cisely determine electrical resistances.
The UNITEST Echometer 3000 can easily and pre-
cisely determine electrical resistances.
The user must check that the circuits are not
live !
live !
" Switch on the UNITEST Echometer 3000 press-
ing the
key.
" Set the measurement range selection to the po-
sition "Ohm”.
" The message ''>1999,9 Ohm'' is displayed.
" Connect the red test lead to the red socket and
" Connect the red test lead to the red socket and
the black test lead to the black socket of the Ohm
range.
range.
" Connect both ends of the resistance to be mea-
sured using the alligator clamps.
The resistance measured is indicated on the display.
If the resistance is below approx. 10 Ohm an
acoustic signal is audible.
If the resistance is below approx. 10 Ohm an
acoustic signal is audible.
Start/Off