NioWatch operation
80
System manual E1
Figure 41: DICOM Optimizer dialog
3. Adjust the slider until there is almost no visible difference 
between the background of the image above the slider and 
the bitmap inside this image.
4. When done, click button Set point 1 of 20.
5. Repeat this procedure until the button is grayed. This 
indicates the last point is set.
6. Click OK to finish.
To reset to the latest calibration:
1. Follow step 2 and 3 of the calibration with Barco LCD sensor 
2. In the Calibration dialog (see step no. 4. in the description of 
the calibration with sensor), check the option Use the latest 
calibration
.