NioWatch operation
82
System manual E1
1. Select the desired test pattern by clicking on the 
corresponding thumbnail or selecting a pattern by name 
from the Test pattern drop-down box.
2.  Click Show to display the pattern.
If you have selected Custom Image, you can select a bitmap 
image (e.g., saved on the hard disk) as test pattern.
3. To hide the test pattern again, click inside the pattern.
Application settings
Calibration tab
Figure 43: Calibration tab
Here you can check the Equalization option, allowing you to 
match all the displays from a multi-head system to the 
luminance of the first calibrated display of the system.
Suppose you work with a system containing 2 displays of the 
same type, and you wish display (2) to have the same output 
luminance as display (1). Therefore first calibrate display (1). 
Next, calibrate display (2) while the Equalization option is 
checked.
The first calibrated display in the system is the reference display.