NioWatch operation
System manual E1
81
3. Click the OK button.
The calibration and the luminance will be reset to the latest 
calibrated position.
Note:
When you start up the system, the calibration is also reset to 
the latest calibrated position. The luminance, however, is 
kept at the latest value.
To view a graph showing the result of the latest calibration:
1. In the Calibration tab dialog, click on Show calibration 
chart.
This button is available only if the display has been 
calibrated with a sensor before.
2. A graph is shown displaying the latest calibration compared 
to the theoretical display function.
To revert to the default factory settings:
1. In the Calibration tab dialog, click on Reset.
2. As a result, the settings are restored to the default factory 
values.
Test patterns
Test Patterns
Figure 42: Test Patterns dialog