Emerson Process Management NGA2000 HFID 用户手册

下载
页码 76
Instruction Manual 
748297-E 
June 2003 
4-6     Maintenance and Troubleshooting 
Rosemount Analytical Inc.    A Division of Emerson Process Management  
Model NGA2000 HFID
The components shown can be replaced without removing burner/thermal block from oven.  Oven not shown for clarity.
Thermal block shown in phantom for clarity. 
Set Screw  
M3X0.5 x 10mm 
903125 
RTD Detector 
657063 
Ignitor Assembly 
657205 
O-Ring 
903736 
O-Ring 
903737
 
Flameout Sensor 
657199 
Temperature Sensor 
657468 
Retainer, Burner Cap 
Burner Cap 
Ignitor 
1.  Refer to Figure 4-4A on page 4-5.  
Remove the four screws on the oven 
cover, remove cover. 
2.  Refer to Figure 4-5 below.  Discon-
nect the ignitor wiring connector, note 
location. 
3.  Using an open-end wrench, unscrew 
the ignitor assembly from the burner.  
Verify that o-ring is also removed. 
4.  Install replacement ignitor and new o-
ring.  Using open-end wrench, snug 
down.  Do not over-tighten!  
5.  Re-attach wiring connector.  
6.  Install oven cover 
 
Flameout Sensor 
1.  Refer to Figure 4-4A on page 4-5.  
Remove the four screws on the oven 
cover, remove cover. 
2.  Refer to Figure 4-5 below.  Discon-
nect the flameout detector wiring con-
nector, note location. 
3.  Lift up the burner cap until flameout 
sensor is accessible.  Using an open-
end wrench, unscrew the flameout 
detector from the burner.  Verify that 
o-ring is also removed. 
4.  Install replacement flameout detector 
and new o-ring.  Using open-end 
wrench, snug down.  Do not over-
tighten! 
 
5.  Re-attach wiring connector.  
6.  Install oven cover.
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 4-5. 
Burner - Sensor, Flameout Detector, RTD Detector and Ignitor