4-8     Maintenance and Troubleshooting 
Rosemount Analytical Inc.    A Division of Emerson Process Management  
Model NGA2000 HFID
Replacing Burner Jets 
Disassemble the burner only if contami-
nants are evident.  Combustion products 
or other contaminates which accumulate 
inside the burner may form electrical 
leakage paths between the collector and 
the burner contact, resulting in noisy read-
ings. 
If the analyzer module is to be operated at 
the highest sensitivity, traces of such con-
taminates can cause erroneous readings.  
For best performance, replace the burner 
jet  follows: 
WARNING 
BURNER  CONTAMINATION 
Do not handle internal parts of the 
burner with bare hands.  All tools used 
for maintenance must be free of con-
taminates. 
 
1.  Remove oven from analyzer module 
per Section 4-2a on page 4-3. 
2.  Remove burner/thermal block from 
oven per 4-2b on page 4-4. 
3.  Remove thermal block from burner 
per Section 4-2c on page 4-7. 
4.  Refer to Figure 4-7A below.  Remove 
screws (2) holding burner cap re-
tainer, remove retainer. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 4-7. 
Burner Disassembly 
 
Burner Cap 
Burner Cap 
Retainer 
Gasket 
656931 
Burner Base 
Burner Base 
Combustion 
Chamber 
B
O-Ring 
904373