Maintenance  and Troubleshooting     4-7 
Model NGA2000 HFID 
c. 
Burner Internal Components 
WARNING 
BURNER CONTAMINATION 
Do not handle internal parts of the 
burner with bare hands.  All tools used 
for maintenance must be free of con-
taminates. 
Disassembly of Burner/Thermal Block 
1.  Remove oven from analyzer module 
per Section 4-2a on page 4-3. 
2.  Remove burner/thermal block from 
oven per Section 4-2b on page 4-4. 
3.  Refer to Figure 4-6 below.  Discon-
nect sample capillary nut at base of 
burner. 
4.  Remove screw securing thermal block 
to burner. 
Carefully pull burner away from thermal 
block. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 4-6. 
Burner/Thermal Block Disassembly 
 
 
 
 
 
 
 
 
 
 
 
Sample capillary
Sample capillary  
input 
Burner 
Thermal Block