4-10     Maintenance and Troubleshooting 
Rosemount Analytical Inc.    A Division of Emerson Process Management  
Model NGA2000 HFID
Installation 
WARNING 
BURNER CONTAMINATION 
Do not handle internal parts of the 
burner with bare hands.  All tools used 
for maintenance must be free of con-
taminates. 
1. Install 
new lower gasket, jet assembly 
and upper gasket into burner base, 
finger-tight jet nut. 
2.  Install new sample jet (with gasket) 
and tighten. 
3.  Tighten jet nut. 
4.  Install air baffle per Figure 4-8 on 
page 4-9. 
Note 
Incorrect installation of air baffle will 
cause ignition failure. 
5.  See Figure 4-7B on page 4-8.  Insert 
new o-ring into burner base. 
6.  Set combustion chamber into burner 
base being careful not to move air 
baffle
7.  See Figure 4-7A on page 4-8.  Insert 
new gasket on combustion chamber, 
install burner cap and burner cap re-
tainer, torque screws to 6 inch lbs.