Emerson Process Management NGA2000 HFID 用户手册

下载
页码 76
Instruction Manual 
748297-E 
June 2003 
Rosemount Analytical Inc.    A Division of Emerson Process Management  
Maintenance  and Troubleshooting     4-9 
Model NGA2000 HFID 
Air Baffle (Ref) 
102260 
See Detail 
Sample Jet 
657005 
Gasket 
102256 
Jet Assembly 
657012 
Gasket 
102273 
Nut, Jet 
657016 
 
 
Air Baffle 
≈  .06
Position with tab over
the air hole
(approximately 1/16” to
the right).
End bent to raise it above the tab a
distance equal to the height of the tab.
5.  Holding burner base, lift burner cap 
off of assembly, set aside. 
6. Remove 
gasket. 
7.  Refer to Figure 4-7B on page 4-8.  
Holding burner base, lift combustion 
chamber off, set aside. 
8.  Refer to Figure 4-8 below.  Lift air baf-
fle out of burner base. 
9.  Remove the sample jet and gasket 
from the bottom of the burner base.  
10.  Remove the jet nut.  Grasp jet as-
sembly and lift out (along with upper 
gasket) of burner base.  Remove bot-
tom gasket. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 4-8. 
Burner Jets