Percent Oxygen Analyzer  
Installation 
 
 
 Teledyne 
Analytical 
Instruments 
25
 
probe or install a longer probe.  This is normally caused by stratification 
of the flue gas.  
 
 
Figure 3-1: Oxygen Probe Mounting  
3.4 Electrical Connections 
Figure 3-1 shows the Model 9060L rear panel.  For detailed 
pinouts, see the wiring/interconnection drawings in the Drawings section 
at the rear of this manual. 
3.4.1 Primary Input Power 
The power cord receptacle, fuse block and Power switch are 
located in the same assembly. A 6-foot, standard AC power cord is 
supplied with the control unit. Insert the female plug end of the power 
cord into the power cord receptacle. 
The power supply allows direct connection to any 110 VAC, 
50/60Hz power source. 220VAC is an optional configuration settable 
through an internal switch in the control unit. The fuse block, to the right 
of the power cord receptacle, accepts two 5x20mm IEC type T, 2 A, 
time-lag (T) fuse. (See Fuse Replacement in chapter 5, Maintenance.) 
The Power switch is located on the right-hand end of the power 
source input receptacle assembly.