Teledyne Percent Oxygen Analyzer 用户手册

下载
页码 51
Installation Model 
9060L 
 
 
Teledyne Analytical Instruments
 
26 
 
Figure 3-2: Model 9060L Rear panel 
3.4.2 Analog Outputs 
There are three DC output signal connectors with screw terminals 
on the panel. There are two wires per output with the polarity noted. See 
Figure 3-3. The outputs are: 
0–10V % Range: 
Voltage rises with increasing oxygen 
concentration, from 0V at 0 percent oxygen to 
10V at full scale percent oxygen. (Full scale = 
100% of programmed range.) 
0–10V Range ID: 
03.33V = Low Range, 06.66V = High Range, 
10V = Air Cal Range. 
4–20 mA % Range:  Current increases with increasing oxygen 
concentration, from 4 mA at 0 percent oxygen to 
20 mA at full scale percent oxygen. (Full scale = 
100% of programmed range.) 
3.4.3 Alarm Relays 
The three alarm-circuit connectors are screw terminals for making 
connections to internal alarm relay contacts. There is one set of contacts 
for each type of alarm. Contacts are Form C, with normally open and 
normally closed contact connections capable of switching up to 1.0 
ampere at 125 VAC into a resistive load (2A for 30 VDC).