ICS-2100 Ion Chromatography System
144
Doc. 065291-01 3/09
4. Remove the EluGen 
cartridge
 VENT
 line 
from the EluGen 
cartridge by 
unscrewing the luer 
lock from the luer 
adapter.
If necessary, use a 
wrench to hold the 
luer adapter in place 
while unscrewing 
the vent line.
Figure 5-35.  EluGen Cartridge Vent Line and Luer Fitting
5. Install the luer cap (P/N 053981), provided with the EluGen cartridge, 
onto the luer adapter. This prevents leaks from the vent opening when 
you turn over the EluGen cartridge 
6. Lift the EluGen cartridge straight up and out of the cartridge holder. 
7. Turn the cartridge upside down and set it in the EluGen cartridge 
service area on top of the ICS-2100 (see 
Figure 5-36.  EluGen Cartridge Service Area
Luer Lock
Luer Adapter
EGC 
Vent
Line
EluGen Cartridge 
Area
Service