Oritron ICS-2100 用户手册

下载
页码 288
5 • Service
Doc. 065291-01 3/09
145
Figure 5-37.  EluGen Cartridge in the Service Area
8. Disconnect the 
EGC IN
 
and 
EGC OUT
 lines 
from the 
INLET
 and 
OUTLET
 ports on the 
EluGen cartridge (see 
Figure 5-38.  Disconnect the EluGen Cartridge Inlet and Outlet Port Lines
9. Follow the instructions in the EluGen cartridge manual, provided on 
the Dionex Reference Library CD-ROM (P/N 053891), to prepare an 
expended EluGen cartridge for disposal or to store a partially used 
EluGen cartridge for later use.
EGC OUT
OUTLET
Port
EGC IN
INLET
Port