5 • Service
Doc. 065291-01 3/09
193
Connect the line labeled 
TO INJ VALVE IN - P/ELUENT OUT
 to the 
EGC CO
3
 Mixer inlet.
Connect the line labeled 
TO PUMP/DAMPER
 to the pulse damper.
4.
Direct the 
WASTE, GAS SEPARATOR
 line from the top of the ICS-
2100 to the rear panel. Snap the line onto one of the tubing clips on 
the rear panel. Connect the line to the gas separator waste tube (see 
5.
5.23.5 Reinstalling the EluGen Cartridge
1. Pull the coiled black 
EGC IN
 and 
EGC OUT
 lines out from inside the 
EGC holder. Disconnect the 
EGC IN
 line from the coupler that 
connects it to the 
EGC OUT
 line.
2. Connect the 
EGC IN 
line
 
to the EluGen 
cartridge 
INLET
 port 
3. Remove the coupler 
and connect the 
EGC 
OUT
 line to the 
EluGen cartridge 
OUTLET
 port.
Figure 5-82.  EluGen Cartridge Inlet and Outlet Port Connections
4. Turn over the EluGen cartridge (fittings facing down). Check for any 
bubbles in the eluent generation chamber, and if necessary, shake and 
tap the EluGen cartridge to remove them.
5. Reinstall the EluGen cartridge in the holder.
6. Reconnect the EluGen cartridge cable and the CR-TC or EPM cable 
(if installed).
Remove the cap from the luer adapter on the EluGen cartridge vent 
opening and re-connect the 
EGC VENT
 line. 
EGC OUT
OUTLET
Port
EGC IN
INLET
Port