5 • Service
Doc. 065291-01 3/09
191
4. Thread the three lines from the EPM (red, yellow, and blue labels) 
through the opening on the side of the new holder (see 
5. Turn the new EGC holder 
NOTE To help retain the
fittings and ferrules
on the lines, remove
the coupler from each
line just before
connecting the line to
a port. 
Figure 5-81.  EGC Holder (upside down)
6. Connect the lines from the EPM (red, yellow, and blue labels) to the 
corresponding lines in the holder. 
7. Disconnect the line labeled 
TO CR-TC/EPM REGEN IN
 (orange label) 
from the 
REGEN IN 
port on the EPM and connect the new 
TO CR-
TC/EPM REGEN IN
 line (orange label) to the
 
port.
Figure 5-80.  TO CR-TC/EPM Lines Outside of Holder
TO CR-TC/EPM 
ELUENT OUT 
(yellow label)
TO CR-TC/EPM
ELUENT IN
(red label)
TO CR-TC/EPM
REGEN OUT
(blue label)