National Instruments NI PXI-5422 用户手册

下载
页码 63
© National Instruments Corporation
47
NI PXI-5422 Calibration Procedure
8.
Take the following voltage measurements at the NI PXI-5422 CH 0 
front panel connector into a high-impedance load:
a.
Call 
niFgen_SetAttributeViInt32
 to set the offset DAC 
value (niFgen property node: Calibration»Offset DAC Value
using the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
channelName
"0"
attributeID
NIFGEN_ATTR_OFFSET_DAC_VALUE
value
50000
b.
Call 
niFgen_Commit
 (niFgen Commit VI) to commit the 
attribute values to the device using the following parameter:
vi: The session handle returned from 
niFgen_InitExtCal
c.
Wait 500 ms for the output to settle.
d.
Use the DMM to measure the voltage output by the device. 
This measurement is 
measurement 0
e.
Call 
niFgen_SetAttributeViInt32
 to set the offset DAC 
value (niFgen property node: Calibration»Offset DAC Value
using the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
channelName
"0"
attributeID
NIFGEN_ATTR_OFFSET_DAC_VALUE
value
15000
f.
Call 
niFgen_Commit
 (niFgen Commit VI) to commit the 
attribute values to the device using the following parameter:
vi: The session handle returned from 
niFgen_InitExtCal
g.
Wait 500 ms for the output to settle.
h.
Use the DMM to measure the voltage output by the device. 
This measurement is 
measurement 1