National Instruments NI PXI-5422 用户手册

下载
页码 63
NI PXI-5422 Calibration Procedure
48
ni.com
9.
Call 
niFgen_CalAdjustMainPathPostAmpGainAndOffset
 
(niFgen Cal Adjust Main Path Post Amp Gain And Offset VI) 
using the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
channelName
"0"
configuration: The Current Configuration value for the current 
iteration from Table 9
mainDACValues: An array containing two elements—the values 
(
0
0
) that you set on the main DAC 
gainDACValues: An array containing one element—the value 
(
2000
) that you set on the gain DAC
offsetDACValues: An array containing two elements—the two 
values (
50000
15000
) that you set on the offset DAC in order
measuredOutputs: An array containing two elements—the 
two output voltages (
measurement 0
measurement 1
) that 
you measured in order
Adjusting the Direct Path Gain
1.
Call 
niFgen_WriteBinary16AnalogStaticValue
 (niFgen Write 
Binary 16 Analog Static Value VI) to set the main DAC value using 
the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
 
channelName
"0"
value
0
2.
Call 
niFgen_SetAttributeViInt32
 to set the gain DAC value 
(niFgen property node: Calibration»Gain DAC Value) using the 
following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
channelName
"0"
attributeID
NIFGEN_ATTR_GAIN_DAC_VALUE
value
2000
3.
Call 
niFgen_SetAttributeViInt32
 to set the offset DAC value 
(niFgen property node: Calibration»Offset DAC Value) using the 
following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
channelName
"0"
attributeID
NIFGEN_ATTR_OFFSET_DAC_VALUE
value
32767