Installation Information
Monitor Installation and Patient Safety   11-5
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
Installation Information
Power 
Source 
Require-
ments
The CMS systems and the Anesthetic Gas Module can be operated from 
an AC source of 100 - 120V 
±
 10%, or 220 - 240V 
±
 10%,
 50 - 60Hz.
Typical input currents for the monochrome and color systems are:
Typical input current for the Anesthetic Gas Module is:
100 - 240V = 1.0 amps
Typical input current for the V24 and V26 is:
90 - 250V = 150VA
Grounding 
the System
To protect the patient and hospital personnel, the cabinet of the installed 
equipment has to be grounded. The equipment is supplied with a 
detachable 3-wire cable which grounds the instrument to the power line 
ground (protective earth) when plugged into an appropriate 3-wire 
receptacle. If a 3-wire receptacle is not available, consult the hospital 
electrician.
Warning
Warning
Do not use a 3-wire to 2-wire adapter.
Monochrome CRT
100 - 120V = 2.2 amps
200 - 240V = 1.2 amps
Color CRT
100 - 120V = 2.6 amps
200 - 240V = 1.4 amps
Color Flatscreen
110 - 120V = 2.0 amps
200 - 240V = 1.0 amps