Installation Information
11-6     Monitor Installation and Patient Safety  
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
If the computer and display module of the M1165/66/75/76A Systems and 
the Anesthetic Gas Module are set up together, connect the grounding 
wire to the equipotential grounding terminal on the display.
If the computer and display module of the M1165/66/75/76A Systems and 
the Anesthetic Gas Module are separated, connect one grounding wire 
to the display, one to the computer module and one to the Anesthetic 
Gas Module.
Equipotential
Grounding
Protection class 1 instruments are already included in the protective 
grounding (protective earth) system of the room by way of grounding 
contacts in the power plug. For internal examinations on the heart or 
the brain, Computer Module and Display Module of the system and the 
Anesthetic Gas Module must have separate connections to the 
equipotential grounding system.
One end of the equipotential grounding cable (potential equalization 
conductor) is connected to the equipotential grounding terminal on the 
instrument´s rear panel and the other end to one point of the 
equipotential grounding system. The equipotential grounding system 
assumes the safety function of the protective grounding conductor if 
ever there is a break in the protective grounding system.
To protect the patient and hospital personnel, the cabinet of the 
installed equipment has to be grounded. The equipment is supplied with 
a detachable 3-wire cable which grounds the instrument to the power 
line ground (protective earth) when plugged into an appropriate 3-wire 
receptacle. If a 3-wire receptacle is not available, consult the hospital 
electrician.
Warning
Warning
Do not use a 3-wire to 2-wire adapter with these instruments
If the computer and display module of the M1165/66/75/76A Systems and 
the Anesthetic Gas Module are set up together, connect the grounding 
wire to the equipotential grounding terminal on the display.