Installation Information
Monitor Installation and Patient Safety   11-7
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
Safety Ground
(Protective
Earth)
 (V24 and V26
only)
Additional Protective Earth (Secondary Ground Wire) Hazardous 
leakage currents can be caused when other equipment is 
interconnected with either monitor.
A secondary ground wire is provided with the CMS, V24 and V26 to 
comply with IEC 601-1-1. This additional protection ensures against 
excessive chassis leakage current in the event of a single fault in the 
health care facility's primary grounding means.
It is recommended that the secondary ground wire be connected to a 
ground source separate from the primary grounding source found in the 
instruments' power source.
This cord should be permanently attached to either monitor mainframe 
rear panel attachment screw.
Check each time before use that the monitor is in perfect working order. 
The cable connecting the patient to the instrument must be free of 
electrolyte.