Controls and Connectors
11-32     Monitor Installation and Patient Safety  
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
M1167/77A
System
The display module (M1095A) cannot be attached to the computer 
module (M1046B). It is designed for mounting on a horizontal or vertical 
rail such as an I.V. pole. For further details, please refer to the CMS 
Service Guide, chapter 2; Installing CMS.
The computer module can be attached to the top or bottom of the 
Anesthetic Gas Module by a locking device. This is a lever which is 
operated from the rear of the monitor, and is attached to the bottom of 
both pieces of equipment.
To assemble the computer module and Anesthetic Gas Module:
Ensure that the locking lever is in the far right position.
Place the Anesthetic Gas Module onto the computer module, 
ensuring that its four feet are in their correct grooves.
Push the locking lever from right to far left. The equipment is now 
secure.
To dismantle the computer module and Anesthetic Gas Module:
Push the locking lever from left to far right.
Lift the Anesthetic Gas Module off the computer module.
M1165/66/
75/76A
 System
The display module (M1094B/92A) can be attached to the computer 
module (M1046A) by a locking device. This is a lever which is operated 
from the rear of the monitor, and is attached to the bottom of the display 
module.
To assemble the computer module and display module:
Ensure that the locking lever is in the far right position.
Place the display module onto the computer module, ensuring that 
its four feet are in their correct grooves.
Push the locking lever from right to far left. The equipment is now 
secure.
To dismantle the computer module and display module:
Push the locking lever from left to far right.
Lift the display module off the computer module.