Controls and Connectors
Monitor Installation and Patient Safety   11-33
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
Lifting the Display Module
The M1092A display module weighs 26 lbs (12kgs)
The M1094A display module weighs 42 lbs (19kgs)
The M1094B display module weighs 43 lbs (19.5kgs)
When carrying the display module, hold it firmly from underneath. For 
safety reasons it is strongly recommended that at least two people 
should lift the display unit. One person on their own should not attempt to 
do so.
Connecting the Anesthetic Gas Module
Caution
If the Anesthetic Gas Module has been stored at sub-zero temperatures, it 
must be left to warm up for a minimum of 4 hours at room temperature, 
before any connection is made to it.
The Anesthetic Gas Module connects to the ACMS over an RS232 cable to 
the Computer Module.
Disposal of the
System
When the system has reached the end of its useful life, it should be 
disposed of in accordance with local regulations for disposal of electronic 
equipment. Plastic parts (the housing for the display, modules and 
computer module) and sheet metal parts (the rear, top and bottom of the 
computer module) can be recycled.
Note—
The FIO
2
 module (M1017A) and the Blood Analysis Module 
(M1022A) contain a Nickel-Cadmium battery, to be disposed of in 
accordance with local requirements.