Controls and Connectors
11-34     Monitor Installation and Patient Safety  
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
The V24 and 
V26 
Connectors
The monitor has several connectors which are shown below.
Do not apply a voltage of more than +12V to any of the pins on the 
defibrillator sync, parameter module rack, Philips monitoring network 
(SDN) or equipotential ground connectors.
The secondary ground complies with IEC 601-1-1 as shown on page 10-4.
 The connectors on the monitor are:
1. AC Power Connector; this connector is used to connect the 
monitor to an AC power source.
Note—
If you have a Philips M1026A Anesthetic Gas Module (AGM), do 
not daisy-chain the AC power through the AGM. This configuration may 
degrade RFI performance. 
2. Secondary Ground Screw; this is a posidrive screw to connect the 
monitor chassis to earth ground per IEC 601-1-1 standard.
3. Equipotential Ground Post, this connects the monitor to the 
equipotential ground system.
AC LINE INPUT
RS232 CONNECTOR
(Rear Housing with
Serial Prefix <3608)
PROTECTIVE
RS232 CONNECTOR
(Rear Housing with
Serial Prefix >=3608)
EQUIPOTENTIAL GROUNDING
DEFIB
CONNECTOR
RACK
CONNECTOR
marge3.tif
EARTH GROUND
PHILIPS MONITORING NETWORK (SDN) 
 SINGLE CONNECTOR