Controls and Connectors
Monitor Installation and Patient Safety   11-35
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
4. Philips monitoring network connector (SDN Interface); this is a 
proprietary Philips system connector, used to input/output 
information to a Philips monitoring network.
Warning
Warning
Connecting the Philips monitoring network (SDN) cable when the 
product is powered on is not supported. Error codes and Philips 
monitoring network (SDN) interface lock-up may occur. Power 
cycling the product will recover the product. No permanent 
damage will result. To prevent unintentional disruption in 
monitoring, be sure the SDN interface cable is properly secured at 
both ends when connecting to the Philips monitoring network 
(SDN).
5. Parameter Module Rack Connector; this connects the parameter 
module rack to the monitor.
6. Defibrillator sync connector; sends an ECG signal out to a 
defibrillator or other external device or receives a marker pulse 
from a defibrillator or other external device. The marker pulse is 
processed with the ECG signal and displayed on the monitor.
Note—
To filter radiated emissions, a core shielding bead (Philips P/N 
9170-1521) must be attached to the defibrillator synchronization cable 
when a defibrillator is connected to the monitor. This filter has no impact 
on the function of the monitor. To attach the core shielding bead:
a. Open the core shielding bead (filter). 
b. Place the bead around the defib sync cable as close to the jack 
as possible. 
c. Close the bead over the cable and snap shut. 
d. Plug the cable into the monitor's defib sync port. The bead 
should be at the monitor end of the cable.