Performance Assurance Checks
13-18     Maintenance  
M
a
int
e
na
nc
e
  
Performing 
the tcpO
2
/
tcpCO
2
 
Module Self-
Test
Auto Check
 - The check is performed automatically when the system is 
switched on, during calibration and continuously when no transducer is 
connected. When the auto check fails, the INOP message “Equipment 
Malfunction'' is given (refer to the Troubleshooting Chapter in the 
service manual for troubleshooting details).
Self-Test 
- The tcpO
2
/tcpCO
2
 module self-test may be performed at any 
time. The test signal lasts about 30 seconds and then the display returns 
to the normal monitoring mode.
TEST - Press 
 
 
RESULT - If the module is functioning correctly, the following numerics 
are displayed without the alarms being activated:
tcpO
2
 60 mmHg (8.0 kPa)
tcpCO
2
 40 mmHg (5.3 kPa)
Caution
Radiated immunity: If operating under conditions according to EN 
60601-1-2 up to 1 V/m electrical field strength, the following deviations in 
the measurements could occur: tcpO
2
 2 mmHg or less, tcpCO
2
 1 mmHg 
or less. It is recommended to avoid the use of electrically radiating 
devices in close proximity to a tcpO
2
/tcpCO
2
 monitor. 
Monitor Setup
Test Signals
Pts
Refernce
Signals