Performance Assurance Checks
Maintenance   13-19
M
a
int
e
na
nc
e
  
Performing 
the CO
2
 
Module Self-
Test
Auto Check 
- None
Self-Test
 - The self-test may be performed at any time. The test signals 
last about 30 seconds, and then the display returns to the normal 
monitoring mode.
TEST - If a transducer is connected to the module, the following 
conditions must be met:
The motor in the transducer is running.
No heater malfunction is reported. 
Press  
 
RESULT - If the module is functioning correctly the values will be 
displayed without the alarms being activated and should read as follows:
ET CO
2
 40 mmHg (6.0kPa)
IMCO
2
 0 mmHg (0.0kPa)
AWRR    25 rpm
Performing 
the 
Temperature 
Module Self-
Test
Auto Check
 - None
Self-Test 
- The Temperature module self-test may be performed at any 
time as the switch network switches to the reference resistors for the 
duration of the check. The system connects the reference resistors to the 
beginning of the processing circuitry, and processes this test value in the 
same way as the patient values are processed. The test signals last about 
30 seconds and then the display returns to the normal monitoring mode.
TEST - Press 
 
 
RESULT - If the module is functioning correctly, the following 
temperature will be displayed without the alarms being activated:
Temp 40 C (+0.1C)