NI 4070/4072 Calibration Procedure
48
ni.com
If you are not performing additional adjustments, refer to one of the 
following sections:
—to verify your new calibration 
coefficients before saving them to the EEPROM
—if you do not want to 
verify the adjustments you have just made
Adjusting AC Voltage (AC- and DC-Coupled) Modes
Note
If you do not use the AC voltage modes for any measurements, or the accuracy of 
these modes is irrelevant, you can skip this section in the calibration procedure and go 
directly to the 
To adjust the AC voltage of the NI 4070/4072, complete the following 
steps:
1.
Reset the calibrator.
2.
Fasten the connectors on one end of the Fluke 5440 cable into the 
appropriate banana plug connectors on the NI 4070/4072, and fasten 
the connectors on the other end of the cable to the appropriate 
calibrator binding posts. Figure 1 shows the correct connections. 
Table 1 lists the cable connections.
3.
Refer to Table 11 for the appropriate calibrator output and parameter 
values as you complete the following steps:
a.
On the calibrator, output the value listed in the Calibrator Output 
column in Table 11 for the current iteration.
b.
Call 
niDMM_CalAdjustGain
 with Mode set to 
NIDMM_VAL_AC_VOLTS
. Set the remaining parameters as shown 
in Table 11 for the current iteration. 
c.
Call 
niDMM_CalAdjustGain
 again, changing Mode to 
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
4.
Repeat step 3 for each of the remaining iterations listed in Table 11.
Table 11.  niDMM_CalAdjustGain Parameters
Iteration
Calibrator Output
niDMM_CalAdjustGain Parameters
Amplitude
f
(kHz)
Mode
Range
(V)
Input Resistance 
Expected 
Value
1
50 mV
1
NIDMM_VAL_AC_VOLTS
0.05
NIDMM_VAL_1_MEGAOHM
0.05
50 mV
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
NIDMM_VAL_1_MEGAOHM
0.05
2
500 mV
1
NIDMM_VAL_AC_VOLTS
0.5
NIDMM_VAL_1_MEGAOHM
0.5
500 mV
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
NIDMM_VAL_1_MEGAOHM
0.5