NI 4070/4072 Calibration Procedure
5.
Refer to Table 12 for the appropriate parameter values as you complete 
the following steps:
a.
Output 0 V on the calibrator.
b.
Call 
niDMM_CalAdjustOffset
 with Mode set to 
NIDMM_VAL_AC_VOLTS
 and the remaining parameters as shown 
in Table 12 for the current iteration. 
c.
Call 
niDMM_CalAdjustOffset
 again, changing Mode to 
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
6.
Repeat step 5 for each of the remaining iterations shown in Table 12.
3
5 V
1
NIDMM_VAL_AC_VOLTS
5
NIDMM_VAL_1_MEGAOHM
5
5 V
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
NIDMM_VAL_1_MEGAOHM
5
4
50 V
1
NIDMM_VAL_AC_VOLTS
50
NIDMM_VAL_1_MEGAOHM
50 
50 V
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
NIDMM_VAL_1_MEGAOHM
50
5
100 V
1
NIDMM_VAL_AC_VOLTS
300
NIDMM_VAL_1_MEGAOHM
100
100 V
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
NIDMM_VAL_1_MEGAOHM
100
Table 12.  niDMM_CalAdjustOffset Parameters
Iteration
niDMM_CalAdjustOffset Parameters
Mode
Range (V)
Input Resistance (
Ω)
1
NIDMM_VAL_AC_VOLTS
0.05
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
NIDMM_VAL_1_MEGAOHM
2
NIDMM_VAL_AC_VOLTS
0.5
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
NIDMM_VAL_1_MEGAOHM
3
NIDMM_VAL_AC_VOLTS
5
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
NIDMM_VAL_1_MEGAOHM
4
NIDMM_VAL_AC_VOLTS
50
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
NIDMM_VAL_1_MEGAOHM
5
NIDMM_VAL_AC_VOLTS
300
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
NIDMM_VAL_1_MEGAOHM
Table 11.  niDMM_CalAdjustGain Parameters (Continued)
Iteration
Calibrator Output
niDMM_CalAdjustGain Parameters
Amplitude
f
(kHz)
Mode
Range
(V)
Input Resistance 
Expected 
Value