National Instruments NI 4070/4072 6 1/2-Digit FlexDMM 用户手册
© National Instruments Corporation
49
NI 4070/4072 Calibration Procedure
5.
Refer to Table 12 for the appropriate parameter values as you complete
the following steps:
the following steps:
a.
Output 0 V on the calibrator.
b.
Call
niDMM_CalAdjustOffset
with Mode set to
NIDMM_VAL_AC_VOLTS
and the remaining parameters as shown
in Table 12 for the current iteration.
c.
Call
niDMM_CalAdjustOffset
again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
6.
Repeat step 5 for each of the remaining iterations shown in Table 12.
3
5 V
1
NIDMM_VAL_AC_VOLTS
5
NIDMM_VAL_1_MEGAOHM
5
5 V
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
NIDMM_VAL_1_MEGAOHM
5
4
50 V
1
NIDMM_VAL_AC_VOLTS
50
NIDMM_VAL_1_MEGAOHM
50
50 V
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
NIDMM_VAL_1_MEGAOHM
50
5
100 V
1
NIDMM_VAL_AC_VOLTS
300
NIDMM_VAL_1_MEGAOHM
100
100 V
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
NIDMM_VAL_1_MEGAOHM
100
Table 12. niDMM_CalAdjustOffset Parameters
Iteration
niDMM_CalAdjustOffset Parameters
Mode
Range (V)
Input Resistance (
Ω)
1
NIDMM_VAL_AC_VOLTS
0.05
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
NIDMM_VAL_1_MEGAOHM
2
NIDMM_VAL_AC_VOLTS
0.5
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
NIDMM_VAL_1_MEGAOHM
3
NIDMM_VAL_AC_VOLTS
5
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
NIDMM_VAL_1_MEGAOHM
4
NIDMM_VAL_AC_VOLTS
50
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
NIDMM_VAL_1_MEGAOHM
5
NIDMM_VAL_AC_VOLTS
300
NIDMM_VAL_1_MEGAOHM
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
NIDMM_VAL_1_MEGAOHM
Table 11. niDMM_CalAdjustGain Parameters (Continued)
Iteration
Calibrator Output
niDMM_CalAdjustGain Parameters
Amplitude
f
(kHz)
Mode
Range
(V)
Input Resistance
Expected
Value