NI 4070/4072 Calibration Procedure
50
ni.com
7.
Refer to Table 13 for the appropriate calibrator outputs and parameter 
values as you complete the following steps:
a.
On the calibrator, output the value listed in the Calibrator Output 
column in Table 13 for the current iteration.
b.
Call 
niDMM_CalAdjustACFilter
 with Mode set to 
NIDMM_VAL_AC_VOLTS
 and the remaining parameters as shown 
in Table 13 for the current iteration. 
Note
The Session parameter remains the same for all instances of this function.
c.
Call 
niDMM_CalAdjustACFilter
 again, changing Mode to 
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
8.
Repeat step 7 for each of the remaining iterations shown in Table 13.
Table 13.  niDMM_CalAdjustACFilter Parameters
Iteration
Calibrator Output
niDMM_CalAdjustACFilter Parameters
Amplitude
Frequency 
(kHz)
Mode
Range 
(V)
Frequency 
(Hz)
1
50 mV 
1
NIDMM_VAL_AC_VOLTS
0.05
1e3
50 mV 
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
1e3
2
50 mV 
5
NIDMM_VAL_AC_VOLTS
0.05
5e3
50 mV 
5
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
5e3
3
50 mV 
20
NIDMM_VAL_AC_VOLTS
0.05
20e3
50 mV 
20
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
20e3
4
50 mV 
50
NIDMM_VAL_AC_VOLTS
0.05
50e3
50 mV 
50
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
50e3
5
50 mV 
100
NIDMM_VAL_AC_VOLTS
0.05
100e3
50 mV 
100
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
100e3
6
50 mV 
200
NIDMM_VAL_AC_VOLTS
0.05
200e3
50 mV 
200
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
200e3
7
50 mV 
300
NIDMM_VAL_AC_VOLTS
0.05
300e3
50 mV 
300
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
300e3
8
50 mV 
500
NIDMM_VAL_AC_VOLTS
0.05
500e3
50 mV 
500
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
500e3
9
500 mV 
1
NIDMM_VAL_AC_VOLTS
0.5
1e3
500 mV 
1
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
1e3